Abstract

Batch treatment with Chelex-100 resin was investigated for preconcentration of trace metals in sea water followed by determination by inductively-coupled plasma atomic emission spectrometry. The preconcentration conditions such as resin weight, stirring time, and amount of ammonium acetate buffer solution were carefully examined for effective multi-element preconcentration from sea water. The resin weight could be decreased to 0.5 g (dry weight) for 1 l of sea water, which was much less than that required in the column method, and a preconcentration factor of 100 was achieved. Al, Ti, V, Mn, Fe, Co, Ni, Cu, Zn, Y, Mo, Cd and Pb in sea water were measured with good precision. The detection limits ranged from 6 to 180 ng l −1. The time required for one sample by the batch method was only 3 h.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.