Abstract

[11C]choline has a weak UV absorption chromophore, and it is challenging to detect less than 1 μg/mL choline using radio-HPLC-UV. In this study, we established an analytical procedure of [11C]choline using the radio-HPLC coupled with the corona-charged aerosol detector. As a result, we achieved more than 100 GBq/μmol (over 0.1 μg/mL of choline) of molar radioactivity at the end of synthesis, over 95% of radiochemical purity, and more than 0.1 μg/mL of 2-dimethyoaminoethanol as a chemical impurity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.