Abstract

In this study we describe the set-up of a new passive integrating system to measure simultaneously 222Rn, 222Rn progeny ( 218Po and 214Po) and 220Rn concentration indoors. It consists of four Makrofol-DE (polycarbonate) circular foils. Two are enclosed within two diffusion chambers—each one with a different filter membrane—to measure 222Rn and 222Rn+ 220Rn. The other two foils are kept in direct contact with air and are electrochemically etched at different conditions to obtain the 222Rn daughters. Theoretical sensitivities of each Makrofol-DE foil are calculated using Monte-Carlo technique. The calculations are performed taking into account: (1) the Bethe–Bloch's expression for the stopping power of heavy charged particles in a medium, (2) the properties and behaviour of 222Rn, 220Rn and their progeny in the open air and within the diffusion chamber and (3) the etching conditions used to visualise α-particles tracks.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.