Abstract

As a technique for calibrating electric-field probes used in standardized SAR (Specific Absorption Rate) assessment, we have studied the technique using the Friis transmission formula in the tissue-equivalent liquid. It is difficult to measure power transmission between two reference antennas in the far-field region due to large attenuation in the liquid. This means that the conventional Friis transmission formula cannot be applied to our measurement so that we developed an extension of this formula that is valid in the near-field region. In this paper, the method of weighted least squares is introduced to reduce the effect of the noise in the measurement system when the gain of the antenna operated in the liquid is determined by the curve-fitting technique. And we examine how to choose the fitting range to reduce the uncertainty of the estimated gain.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.