Abstract We describe a method for simultaneously measuring concentrations of coproporphyrin, zinc protoporphyrin IX, and protoporphyrin IX in whole blood by liquid chromatography, with use of reversed-phase ion-pair system, fluorometric detection, and internal standardization. Each analysis requires 10 microL of whole blood and 15 min total analysis time. Analytical recovery ranged from 84 to 92%, day-to-day precision (CV) from 5 to 12%. Uroporphyrin, though not studied in detail, can also be detected by this method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.