In order to improve the resolving power (RP) and signal-to-noise ratio (SNR) of ion mobility spectrometry (IMS) simultaneously, a modified Hadamard transform-inverse ion mobility spectrometry (MHT-IIMS) technique was developed. In this novel technique, a series of isolating codes were appended to each element of the pseudo random binary sequence (PRBS), and then the modified modulation sequence was formed and used to control the ion gate of the inverse IMS (IIMS). Experimental results demonstrate that the MHT-IIMS technique can significantly enhance the resolving power and SNR simultaneously by measuring the spectra of reaction ions. Furthermore, the gas sample CCl4 and CHCl3 were measured for evaluating the capability of detecting those samples which have single and multiple product ions. The results show that this novel technique is able to simultaneously improve the resolving power and SNR notablely for the real sample detection without any significant instrumental changes. Graphical Abstract ᅟ.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the American Society for Mass Spectrometry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.