Abstract

A TLC-densitometric method has been developed for simultaneous identification and quantitative determination of amikacin, gentamicin, kanamycin, neomycin, netilmicin, and tobramycin. This separation of antibiotics was achieved on silica gel TLC plates without a fluorescent indicator and with methanol-25% ammonia-chloroform (3 + 2 + 1, v/v/v) as the mobile phase. The densitometric measurements were made at 500 nm after detection with a 0.2% ninhydrin solution in ethanol. Under these conditions, good separation of the chosen aminoglycosides was obtained. The method is distinguished by high sensitivity, with the LOD from 0.25 microg for amikacin to 1.00 microg for gentamicin and the LOQ from 0.5 microg for amikacin to 1.65 microg for gentamicin, and a wide linearity range 0.75-6.25 microg/spot for amikacin and netilmicin and 1.5-12.50 microg/spot for other antibiotics. The precision of the determination was very good; RSD varied in the range 0.3-0.6%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.