Abstract
The simultaneous identification and quantitation of 15 benzodiazepines and selected metabolites in postmortem blood, serum, or liver homogenate is described. The assay involves extraction with diethylether, followed by an acid clean-up step of the ether. Chromatographic separation was achieved on a Nova-Pak phenyl 18 column using ultraviolet detection at 240 nm. A gradient HPLC system was developed to improve separation of nitro-reduction metabolites from the solvent front and endogenous peaks. The mobile phases consisted of a gradient from 15 to 28% acetonitrile in 40 mM potassium phosphate buffer. Within-run and day-to-day precision were generally 10-15%. The method described is sensitive and reproducible for the analysis of benzodiazepine concentrations in postmortem tissues.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.