Abstract

Abstract In the present work, an easy solid phase extraction method using alumina modified with polyethylenimine as a new adsorbent was applied to the simultaneous extraction of copper, silver, and palladium ions prior to their determination with electrothermal atomic absorption spectrometry. The analytical procedure involved the complex formation of these cations with polyethylenimine as a chelating agent in buffer media of pH 7.0. Under the optimum conditions, a preconcentration factor of 200, 150, and 200, precision of ±5.4%, ±4.7%, and ±5.2% and linear calibration ranges of 15.0–140, 4.0–93, and 7.5–125 ng/L (in original solution) for Cu, Ag, and Pd were obtained, respectively. Also detection limits of 3.9, 1.1, and 2.0 ng/L were obtained for Cu, Ag, and Pd, respectively. The proposed method was applied to the determination of copper, silver, and palladium in some real samples with satisfactory results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.