Abstract

: The bio-analytical method was developed and validated for simultaneous detection and quantification of paclitaxel (PAC) and erlotinib (ERL) in plasma samples. The sample preparation process was accomplished by liquid -liquid extraction technique. The dried and reconstituted samples were subjected to chromatography on Discovery -C18 (50 × 4.6 × 5µm) column and a mobile phase, composed of a mixture of 0.1% formic acid in water: acetonitrile (70:30, v/v), in isocratic mode at a flow rate of 0.6 mL/min. Liquid chromatography coupled to tandem mass spectrometry detection in positive ion mode was selected to provide optimal selectivity and sensitivity. The mass transitions of erlotinib, erlotinib13C6, Paclitaxel and docetaxel were m/z 394.5→278.4, m/z 400.4→284.5, m/z 876.6→308.4 and m/z 830.0→304.0 respectively. The linearity in the calibration curves were obtained in the concentration range of 3.6 -1006.7 ng/ml (r ≥ 0.99) for erlotinib and 5.3 -1500.0 ng/mL for paclitaxel with a LLOQ (lower limit of quantification) of 3.6 and 5.3ng/ml respectively. The run time was achieved in 2.5 minutes only, for all the analytes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.