Abstract

Simultaneous electrokinetic and hydrodynamic injection (SEHI) of organic cations (tricyclic antidepressant and beta blocker drugs) with on-line sample concentration using micelle to solvent stacking (MSS) was studied in micellar electrokinetic chromatography. Compared to conventional injection, >300-fold improvements in signals were obtained by hydrodynamic injection. However, with SEHI the amount of sample ions introduced into the capillary was increased which afforded a higher gain of up to 4000-fold without compromise to separation efficiency. The electrokinetic injection at negative polarity (anode at the detector end) introduced the micelle bound analytes. The hydrodynamic injection also maintained the MSS boundary inside the capillary. The stability of the MSS boundary affected SEHI where mild conditions that were low voltage as well as pressure injection were desired. The limits of detection were in the range from 0.6–4.2ngmL−1. A strategy for optimization was described and the method was applied to the ngmL−1 analysis of spiked wastewater after simple dilution and centrifugation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.