Abstract

Assays of copper and zinc in brass samples were performed by semi-mercury free potentiometric stripping analysis (S-MF PSA) using a thin-film mercury covered glassy-carbon working electrode and dissolved oxygen as oxidizing agent during the stripping step. The stripping peak transients were resolved by chemometrics, which enabled simultaneous determination of both the copper and the zinc concentrations, thereby eliminating the conventional necessary pretreatment of the sample solution, such as initial addition of Ga(III) or solvent extraction of copper. The brass samples were diluted by factors in the range 2 · 104– 5 · 105 which resulted in quantification of the copper and of zinc contents comparable to the specified values within 10%. On the basis of the chemometric treatment, an empirical expression is deduced relating the stripping time to the recorded potential.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.