Abstract

A validated adsorptive cathodic stripping voltammetry method is described for simultaneous determination of Al(III), Cu(II) and Cd(II) in water samples. In acetate buffer (pH 5) containing 10 μM oxine, these metal ions were determined as oxine complexes following adsorptive accumulation onto the HMDE at −0.05 V versus Ag/AgCl/KCls. The best signal to noise ratio was obtained using a square wave of scan increment 10 mV, frequency 120 Hz, and pulse-amplitude 25 mV. Limits of detection as low as 0.020 μg L−1 Al(III), 0.012 μg L−1 Cu(II) and 0.028 μg L−1 Cd(II) were achieved. Interference due to various cations (K(I), Na(I), Mg(II), Ca(II), Mn(II), Fe(III), Bi(III), Sb(III), Se(IV), Pb(II), Zn(II), Ni(II), Co(II)), anions (Cl−, NO3−, SO42−, PO43−) and ascorbic acid was minimal as the measured signals change by 4% at the maximum. The stripping voltammetry method was successfully applied for simultaneous determination of Al(III), Cu(II) and Cd(II) in tap and natural bottled water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.