Abstract
The design, development, and testing of a flow-injection analysis (FIA) manifold for the simultaneous determination of two species in milk samples is reported. Calcium is determined potentiometrically, and, following on-line sample treatment, spectrophotometric analysis of phosphorus is performed. Results obtained are consistent with manufacturers’ estimates and, in the case of calcium, with atomic spectrophotometric measurements. The combination FIA system is a rapid, low-cost, highly reproducible method that uses small sample volumes and could be used to provide on-site near-real-time measurements. © 1998 John Wiley & Sons, Inc. Lab Robotics and Automation 10: 235–238, 1998
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.