Abstract

A highly sensitive, precise, and accurate reversed-phase high performance liquid-chromatography/electrochemical detection method for simultaneous determination of the endogenous free α-lipoic acid and dihydrolipoic acid in biological matrices is presented. The two analytes are extracted from samples with acetonitrile-10% m-phosphoric acid solution(aqueous) (50:50 v/v). To determine the total lipoic acid, samples are treated with tris(2-carboxyethyl)phosphine solution in phosphate buffer: pH 2.5 with 85% o-phosphoric acid prior to deproteination. The two analytes are separated on a C18 (150 × 4.6 mm, 5 μm) analytical column using acetonitrile-50 mM phosphate buffer: pH 2.5 with 85% o-phosphoric acid (35:65 v/v) as the isocratic mobile phase pumped at a flow rate of 2.0 ml/min at the column oven temperature of 35 °C. The column eluents are monitored at a potential of 0.9 V. These analytes are efficiently resolved in <7 min.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.