Abstract
A method is described for simultaneous determination of semivolatile organic compounds (SVOCs) in indoor air by gas chromatography–mass spectrometry (GC–MS). The selected 73 SVOCs were collected using combined adsorbents (quartz fiber filter disk and Empore disk) for 24 h at a 5.0 l/min flow rate. The SVOCs collected were extracted with acetone, concentrated, then analyzed by an internal standard method. Forty compounds (19 plasticizers and flame retardants; 19 insecticides; 1 synergist; and 1 fungicide) among the target SVOCs were determined accurately and precisely. The method of detection limits for these compounds were approximately 0.5 ng/m 3 for most of the SVOCs. The collected SVOC samples could be stored for up to 1 month at 4 °C in the refrigerator.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.