Abstract
A flow injection procedure was developed for the simultaneous determination of two UV filters (oxybenzone and 2-ethylhexyl 4-methoxycinnamate) in sunscreen formulations, based on the isodifferential approach. The use of the second derivative spectra allowed a selective determination of the analytes free from matrix interferences and without pretreatment of samples. Standard solutions of the analytes were analyzed by the proposed procedure in order to select the appropriate wavelengths to solve the mixture. After this, the procedure was validated using commercial sunscreen formulations the concentrations of which were determined by an HPLC procedure. The two procedures gave comparable results. The sensitivity for oxybenzone determination by the proposed method was (9.2±0.1)×10 −5 ml μg −1, and the limit of instrumental detection was 0.5 μg ml −1. Determination of 2-ethylhexyl 4-methoxycinnamate had a sensitivity of (4.6±0.1)×10 −5 ml μg −1 and a limit of detection of 2.1 μg ml −1. The proposed flow injection procedure allowed on-line dilution of samples, and a sampling frequency of 360 h −1. The accuracy, speed and selectivity of the procedure make it suitable for this type of formulations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.