Abstract
Urine samples are treated with concentrated nitric acid and potassium permanganate ar 70°C for 10 min prior to injection. The flow electrode system consists of a 10-μm diameter gold fibre working electrode, a glassy carbon reference electrode and a platinum counter electrode. In the fully automated constant-current stripping procedure, the gold fibre is first covered with a fresh gold film after which the sample is electrolyzed for 1 min prior to stripping in 0.1 M hydrochloric acid with a current of 0.1μA. The procedure is repeated on a spiked sample after which the sample analyte concentrations are evaluated and presented digitally and graphically on a printer/plotter. The results obtained for bismuth, copper and mercury in a urine reference sample were 36.9, 39.7 and 47.7 μg l −1 with standard deviations ( n=10) of 3.2, 4.2 and 2.1, respectively. The certified values for copper and mercury were 45 and 51 μg l −1; no certified value was available for bismuth.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.