Abstract

A gradient HPLC method coupled with diode array detection was developed and fully validated for the analysis of impurities in ropinirole using a Kromasil® C8 100 A (250 × 4.6 mm, 5 μm) column with a flow rate 1.0 mL min−1 and detection at 250 nm. The mobile phase component A consisted of a mixture of 19.6 mM aqueous potassium dihydrogen phosphate–acetonitrile (98:2 v/v), pH adjusted to 7.0 with triethylamine and the mobile phase component B consisted of acetonitrile. The method was validated in terms of linearity, sensitivity, precision, accuracy and stability. The calibration curves for ropinirole and its five impurities showed good linearity (r > 0.998) within the calibration ranges tested. The intra- and inter-day RSD values were <3.9 %, while the relative percentage error E r was <5.8 % for all compounds. Accelerated stability studies performed under various stress conditions including oxidation, hydrolysis and UV light irradiation at 254 nm proved the selectivity of the procedure. Long-term stability studies performed on blistered tablets and under various conditions of heat and humidity indicate the presence of four of the studied impurities in less than 0.07 %. The method was applied successfully to the detection and determination of ropinirole impurities in pharmaceutical formulations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.