Abstract
A rapid and sensitive liquid chromatography-tandem mass spectrometry (LC-MS/MS) method has been developed and validated for the simultaneous estimation of hydrochlorothiazide, quinapril and its metabolite quinaprilat in human plasma. After solid phase extraction (SPE), the analytes and IS were chromatographed on a hypurity C8 (100 mm × 2.1 mm i.d., 5 μm particle size) column using 2 μL injection volume with a run time of 2.8 min. An isocratic mobile phase consisting of 0.5% (v/v) formic acid:acetonitrile (25:75, v/v) was used to separate all these drugs. The precursor and product ions of these drugs were monitored on a triple quadrupole mass spectrometer, operating in the multiple reaction monitoring mode (MRM) without polarity switch. The proposed method was validated over the range of 5–500 ng/mL for hydrochlorothiazide method and 5–1500 ng/mL for quinapril and quinaprilat. Inter-batch and intra-batch precision (coefficient of variation – % CV) across five validation runs lower limit of quantitation (LLOQ), lower quality control (LQC), middle quality control (MQC), higher quality control (HQC) and upper limit of quantitation (ULOQ) was less than 15. The accuracy determined at these levels was within ±13% in terms of relative percentage error.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.