Abstract
A high-performance liquid chromatography (HPLC)-ultraviolet detection method, combined with solid-phase extraction (SPE), was developed for the determination of five nitroaniline and dinitroaniline isomers including 2-nitroaniline, 3-nitroaniline, 4-nitroaniline, 2,4-dinitroaniline and 2,6-dinitroaniline in wastewater samples. Extraction of the five isomers was carried out with a hydrophile–lipophile balance cartridge, the Oasis HLB. The cartridge was washed by a mixed aqueous solution containing 10% (v/v) acetonitrile and 10% (v/v) ethyl acetate before the five isomers were eluted by a mixture of methanol and acetic acid. Separation of the five isomers was achieved by using an Agilent TC-C 18 column at 30 °C, and using a mixture of acetonitrile/water 30/70 (v/v) as mobile phase under an isocratic condition at a flow rate of 1.0 mL/min. The analytes were detected by a UV detector at a wavelength of 225 nm. Recoveries of the five isomers in the spiked sewage sample were between 84.6% and 94.0% with a relative standard deviation of less than 4.7%. The limits of quantification (LOQ) determined in a spiked sewage sample of 500 mL were 2.0 × 10 −9 M for 2-nitroaniline, 3-nitroaniline and 2,6-dinitroaniline, and 4.5 × 10 −9 M for 4-nitroaniline and 2,4-dinitroaniline. The proposed method was applied to determine the five isomers in real samples of acidic wastewater and printing and dyeing wastewater.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.