Abstract
In this study, we developed an ultra-performance liquid chromatography-electrospray tandem quadrupole mass spectrometry (UHPLC-ESI-MS/MS) method to simultaneously determine Picroside-I, Picroside-II, Picroside-III, minecoside, and sweroside in rat plasma. The chromatographic column was an ACQUITY UHPLC® BEH Amide Column (2.1 × 100 mm, 1.7 µm; Waters, MA, USA), column temperature 40 °C. The mobile phase was 0.1% formic acid aqueous solution-0.1% formic acid acetonitrile solution. The flow rate was 0.4 mL/min. Multiple reaction monitoring (MRM) and negative ion modes were adopted. The results showed that the calibration curves of five compounds in plasma showed good linearity (r > 0.9911) over the studied dose range. The lower limits of quantification (LLOQ) for Picroside-I, Picroside-II, Picroside-III, minecoside, and sweroside were 6.876, 5.193, 5.040, 1.260, and 4.527 ng/mL, respectively. The intra-day and inter-day precision were <15%. The matrix effects ranged from 95.77 to 101.9%. The Tmax were 1.1 ± 0.2, 1.1 ± 0.1, 0.8 ± 0.1, 1.0 ± 0.2, and 2.1 ± 0.1 h. This study will be useful in understanding the behavior of drugs in the body and the body's effect on drugs. It also offers theoretical underpinnings and highlights the importance of clinical applications and creating novel drugs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.