Abstract

A simple approach based on cyclic voltammetry (CV) was developed for the simultaneous determination of dopamine (DA) and uric acid (UA) in the presence of ascorbic acid (AA) using a modified glassy carbon paste electrode (GCPE). In the present study, analytical parameters were optimized and electrochemical performance of modified electrode was investigated. The calibration curves were obtained over the range of 15.97-157.98 μmol L-1 dopamine and 15.97-195.35 μmol L-1 uric acid. Detection limits of 12.38 μmol L-1 dopamine and 3 μmol L-1 uric acid were also obtained at pH 7.0. The modified electrode was used for practical application for the detection of dopamine and uric acid in the real samples of uric acid and dopamine injection, human serum and urine samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.