Abstract

A simple and selective reversed-phase high-performance liquid chromatography method was developed for the estimation of cromolyn sodium (CRM) with either oxymetazoline hydrochloride (OMZ) or xylometazoline hydrochloride (XMZ) in their binary mixtures. The method is based on the simultaneous separation of each drug in a reversed-phase Waters symmetry® C18 column (250 mm×4.6 mm intradermally, 5-µm particle size) at 25°C. Elution was performed with a mobile phase consisting of methanol : 0.1 M phosphate buffer (60:40, v/v, pH 4.0). Quantitation was achieved with ultraviolet detection at 220 nm. The method could determine the three drugs, with linearity, in the range of 2.0-100.0 µg/mL for CRM and 0.8-8.0 µg/mL for OMZ and for XMZ. Aspirin was used as internal standard. Optimization of the separation in terms of mobile phase composition is critical to the method development, which is discussed in detail. The suggested procedure was successfully applied to the analysis of the studied drugs in their nasal preparations. Statistical evaluation of the data obtained by the proposed and comparison methods revealed good accuracy of the proposed method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.