Abstract

A gas chromatography–mass spectrometry method was developed and validated for the simultaneous automated solid-phase extraction and quantification of cotinine and trans-3-hydroxycotinine in human urine. Good linearity was observed over the concentration ranges studied (R2 > 0.99). The limit of quantification was 10 ng/mL for both analytes. The limits of detection were 0.06 ng/mL for cotinine (COT) and 0.02 ng/mL for trans-3-hydroxycotinine (OH-COT). Accuracy for COT ranged from 0.98 to 5.28% and the precision ranged from 1.24 to 8.78%. Accuracy for OH-COT ranged from −2.66 to 3.72% and the precision ranged from 3.15 to 7.07%. Mean recoveries for cotinine and trans-3-hydroxycotinine ranged from 77.7 to 89.1%, and from 75.4 to 90.2%, respectively. This analytical method for the simultaneous measurement of cotinine and trans-3-hydroxycotinine in urine will be used to monitor tobacco smoking in pregnant women and will permit the usefulness of trans-3-hydroxycotinine as a specific biomarker of tobacco exposure to be determined. © 2014 The Authors. Biomedical Chromatography published by John Wiley & Sons Ltd.