Abstract

This study demonstrates the separation of active ingredients in acne formulations (salicylic acid, cloramphenicol and resorcinol in presence of azulene) by capillary zone electrophoresis. Factors affecting their separations were the buffer pH and concentration, applied voltage, sample preparation, and presence of additives. Optimun results were obtained with a 50 mM sodium tetraborate–50 mM sodium phosphate, pH 9.0. The carrier electrolyte gave baseline separation with good resolution, short migration times (<6 min), great reproducibility and accuracy. Calibration plots were linear over at least three orders of magnitude of analyte concentrations, the lower limits of detection being within the range 0.39–1.25 μg ml −1. The procedure was fast and reliable and commercial pharmaceuticals could be analysed without prior sample clean-up procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.