Abstract
Simultaneous determination of carboxylic acids and inorganic anions was investigated by photometric ion chromatography with phthalate solution as an eluent. The retention time and the peak area of carboxylic acids changed remarkably with eluent pH. The optimum conditions at present were as follows : analytical column, OYOBUNKO ASA-4000 (4.6 mm i.d.×25 cm) ; eluent, 5×10-4 M disodium phthalate ; flow rate, 2 ml/min ; column temperature, 40°C ; detection wavelength, 240 nm ; injection volume, 100 μl. The detection limits for several carboxylic acids as 3 times noise were 0.05-0.07 mg/l in sample solution. Their calibration curves showed straight lines by the peak area method. The present method has an advantage to determine simultaneously not only carboxylic acids but also inorganic anions. Under the condition described above, both carboxylic acids and inorganic anions in several liquid foods were easily determined by direct injection of the sample prepared with only dilution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.