Abstract

This paper describes the simultaneous determination of Ba, Co, Fe, and Ni in nuts by high-resolution continuum source atomic absorption spectrometry after extraction induced by solid-oil-water emulsion breaking. Extraction yields ranged from 94.9 for Ba to 109.8% for Fe. Simultaneous measurements were carried out at secondary lines of Ba, Co, Fe, and Ni. The limits of detection and quantification were, respectively, 3.819 and 1.146 mg L−1 for Ba, 2.274 and 7.421 µg L−1 for Co, 0.095 and 0.285 mg L−1 for Fe, and 2.138 and 6.614 µg L−1 for Ni. The precision ranged from 3.1 to 4.2%, 1.5 to 8.0%, 1.6 to 6.6%, and 0.4 to 6.1% for Ba, Co, Fe and Ni, respectively. The method accuracy was assessed by recovery tests and comparison of the results obtained by the proposed extraction method with those obtained after acid digestion. Recoveries ranged from 93.5 for Ni to 104.5% for Co.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.