Abstract

We developed a method to analyze atmospheric SO(x) (particulate SO(4)(2-)+ gaseous SO(2)) and NO(x) (NO + NO(2)) simultaneously using a battery-operated portable filter pack sampler. NO(x) determination using a filter pack method is new. SO(x) and NO(x) were collected on a Na(2)CO(3) filter and PTIO (2-phenyl-4,4,5,5-tetramethylimidazoline-3-oxide-1-oxyl) + TEA (triethanolamine) filters (6 piled sheets), respectively. Aqueous solutions were then used to extract pollutants trapped by the filters and the resulting extracts were pre-cleaned (e.g. elimination of PTIO) and analyzed for sulfate and nitrite by ion chromatography. Recoveries of SO(2) and NO(x) from standard pollutant gases and consistency of the field data with those from other instrumental methods were examined to evaluate our method. SO(x) and NO(x) could be analyzed accurately with determination limits of 0.2 ppbv and 1.0 ppbv (as daily average concentrations), respectively. The sampler can determine SO(x) and NO(x) concentrations at mountainous or remote sites without needing an electric power supply.