Abstract

Abstract This article deals with the design of an unsegmented-flow injection manifold for the simultaneous determination of adrenaline and noradrenaline two structurally related compounds with overlapping spectra. An FIA manifold is proposed for the simultaneous determination in which the sample solution is directly injected into a carrier-reagent stream of aqueous NaOH. The selected wavelengths (first derivative) were 394 and 342 nm, for noradrenaline and adrenaline, respectively with an integration time of 0.4 s. The calibration graphs are linear over the range 2.0–30 ppm for both drugs. The method is applied to different synthetic mixtures of both drugs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.