Abstract
AbstractA new method has been described to determine both benzodiazepines (six) and tricyclic antidepressants (four) simultaneously in saliva by HPLC with a UV detector set at 240 nm using cholchicine as the internal standard. A careful specific sequential solid‐phase elution was optimized and performed to elute benzodiazepines using a mixture of methanol‐acetonitrile (1:1 v/v) followed by the elution of tricyclic antidepressants with methanol. Separation of the compounds was performed on a Kromasil column (250 × 4 mm, 5 μm) by a gradient eluents consisting of 0.05 M CH3COONH4‐acetonitrile‐methanol (55:15:30 v/v/v). The results were linear for both benzodiazepines and tricyclic antidepressants up to 20 ng μL‐1 with the correlation coefficients greater than 0.998. The sensitivity limits, LOD and LOQ were 0.08‐0.34 ng μL‐1 and 0.28‐1.13 ng μL‐1, respectively. The method is simple, fast and reliable with good specificity and sensitivity, will be suitable for use in a clinical setting, where there is a concomitant use of 1,4‐benzodiazepines and tricyclic antidepressants.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.