Abstract

Objective: To describe for the determination of contents of metabolites of benzene compounds in urine sample by high performance liquid chromatography. Methods: After acidification with hydrochloric acid, metabolites in urine were first extracted by acetonitrile and isopropanol (V∶V, 9∶1) with excessive sodium chloride, then gradient separated on a C18 column and then determined by DAD detector. Results: There were good linear relationship between peak areas and injection quality in range of 2.00-100 mg/L (r>0.999). The detection limit and quantitative limit of this method were 4.15-70.7 μg/L and 13.8-235 μg/L respectively. The precision for the analysis of urine was1.78%-8.23% (n =6). The average recovery of metabolites was 85.4%-105.5% at thee spiked levels in the range of 2.00-100 mg/L. Conclusion: The accuracy and reproducibility obtained make this method useful for the biological monitoring of occupational exposure to toluene, xylene, styrene and ethylbenzene.