Abstract

ABSTRACT In the present study, a simple and fast method for simultaneous detection of zearalenone, citrinin, and ochratoxin A utilising capillary zone electrophoresis with an ultraviolet detector was developed. The optimised approach was validated and applied using pepper samples. The proposed method yielded satisfactory linearity between the signal and the mycotoxin concentration in the range of 1.5–150 μg/kg for zearalenone, 4.5–150 μg/kg for citrinin, and 0.8–150 μg/kg for ochratoxin A. The limits of detection for these mycotoxins ranged from 0.3 to 1.5 μg/kg. The corresponding intra- and inter-day precisions were less than 3.5 % and 4.1 %, respectively. Moreover, the matrix effect was also assessed and the result was compared using the capillary zone electrophoresis and high-performance liquid chromatography methods. The developed approach could be used for simultaneous detection of zearalenone, citrinin, and ochratoxin A in pepper.