Abstract

A LC/MS method for the simultaneous determination of both type A and type B trichothecenes by using an electrospray ionization (ESI) interface in the positive ionization mode with a single quadrupole analyzer is described. In order to enhance the ionization of both groups of trichothecenes, the sodium ion was used as cationization agent by adding sodium chloride to the eluent. All LC/MS parameters were optimized. The newly developed LC/ESI-MS method was applied to the analysis of a wheat reference material and cereal-based foods and feeds.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.