Abstract

A two dimensional ion chromatography (2D-IC) method was improved for the simultaneous determination of six cations and seven anions using one-valve switching after injection and conductivity detector. The trap conditions and valve-switch timing were optimized. The method has good accuracy and sensitivity for the simultaneous separation of inorganic anions and cations with a mean correlation coefficient of >0.99, repeatability of 0.60%–3.0% for eight replicates, reproducibility of 0.78%–3.1%, and spiked recovery rate of 88.7%–110%. The method was successfully applied to the analysis of inorganic anions and cations in four kinds of mineral water and can completely satisfy the demand of water quality monitoring.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.