Abstract
This study investigated the audiometric and sound localisation results in patients with conductive hearing loss after bilateral Bonebridge implantation. Eight patients with congenital microtia and atresia supplied with bilateral Bonebridge devices were enrolled in this study. Hearing tests and sound localisation were tested under unaided, unilateral and bilateral aided conditions. Mean functional gain was higher with a bilateral fitting than with a unilateral fitting, especially at 1.0-4.0 kHz (p < 0.05, both). The improvement in speech reception threshold in noise with a bilateral fitting was a 2.3 dB higher signal-to-noise ratio compared with unilateral fitting (p < 0.05). Bilateral fitting had better sound localisation than unilateral fitting (p <0.001). Four participants who attended follow up showed improved sound localisation ability after one year. Patients demonstrated better hearing threshold, speech reception thresholds in noise and directional hearing with bilateral Bonebridge devices than with a unilateral Bonebridge device. Sound localisation ability with bilateral Bonebridge devices can be improved through long-term training.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.