Abstract

A normal-phase high performance liquid chromatography method for the simultaneous determination of tert-butylhydroquinone, tert-butylquinone, butylated hydroxytoluene, 2-tert-butyl-4-hydroxyanisole, 3-tert-butyl-4-hydroxyanisole, α-tocopherol, γ-tocopherol, and δ-tocopherol in edible oils was investigated. A silica column was used to separate the analytes with the gradient elution. An ultraviolet–visible detector was set at dual wavelengths mode (280 and 310nm). The column temperature was 30°C. The analytes were directly extracted with methanol. Results showed that the normal-phase high performance liquid chromatography method performed well with wide liner ranges (0.10∼500.00μg/mL, R2>0.9998), low limits of detection and quantitation (below 0.40 and 1.21μg/mL, respectively), and good recoveries (81.38∼102.34% in soybean oils and 83.03∼100.79% in lard, respectively). The reduction of tert-butylquinone caused by the reverse-phase high performance liquid chromatography during the injection was avoided with the current normal-phase method. The two isomers of butylated hydroxyanisole can also be separated with good resolution.