Abstract
The neutron flux monitor (NFM) is one of the most important diagnostic systems for ITER. Wide-range measuring algorithm (WRMA) is the core algorithm in the NFM system, which deals with the key task of neutron flux measurement. In this paper, the principle and implementation of WRMA, including counting and Campbelling algorithms, are introduced in detail, with error sources of the two algorithms analyzed. In order to study the performance of WRMA, we established a simulation system for neutron signal processing using MATLAB. According to the principle of neutron pulse distribution, the digital waveforms at different neutron flux levels were simulated as inputs to the WRMA module. The variation of measuring error was studied by comparing the counting and Campbelling results with actual input counting rate. In addition, the effects of different neutron pulse widths on the results of the algorithm were simulated. A preliminary experiment at HL-2A was carried out to validate the algorithm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.