Abstract

The optimized size of a single-channel surface radio frequency (RF) coil for mouse body images in a 9.4 T magnetic resonance imaging (MRI) system was determined via electromagnetic-field analysis of the signal depth according to the size of a single-channel coil. The single-channel surface RF coils used in electromagnetic field simulations were configured to operate in transmission/reception mode at a frequency of 9.4 T–400 MHz. Computational analysis using the finite-difference time-domain method was used to assess the single-channel surface RF coil by comparing single-channel surface RF coils of varying sizes in terms of |B1|-, |B1+|-, |B1−|- and |E|-field distribution. RF safety for the prevention of burn injuries to small animals was assessed using an analysis of the specific absorption rate. A single-channel surface RF coil with a 20 mm diameter provided optimal B1-field distribution and RF safety, thus confirming that single-channel surface RF coils with ≥25 mm diameter could not provide typical B1-field distribution. A single-channel surface RF coil with a 20 mm diameter for mouse body imaging at 9.4 T MRI was recommended to preserve the characteristics of single-channel surface RF coils, and ensured that RF signals were applied correctly to the target point within RF safety guidelines.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.