Abstract
A De-Pangher-type long counter was designed for neutron measurements in standard neutron fields based on the results of simulations using the MCNP6 code at the National Metrology Institute of Japan, the National Institute of Advanced Industrial Science and Technology. The effects of six parameters in the design on the energy response of the long counter were investigated. The energy response was then quantitatively evaluated for eight design candidates selected from the investigation. The calculation results show that these candidates have a flatter energy response from 100eV to 10MeV compared to that of the current long counter. These candidates also reduce the difference between the average energy response from 10eV to 1MeV and that from 10 to 20MeV.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.