Abstract
Simulation studies were performed on calculation of the detection efficiency for a phoswich detector with background rejection capability for continuous blood sampling system. The detector consists of a plastic scintillator, a BGO scintillator, and a photomultiplier tube (PMT). The pulse shape information is used to distinguish whether the detected event is a true event or a background one. First, we calculated the transmission efficiency of the positron in the tube. And we calculated the detection efficiency of positrons detected by the plastic scintillator. We also calculated the detection efficiency of 511 keV gamma photons by the BGO.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.