Abstract

As a new energy source for atrial fibrillation ablation, electric pulse ablation has higher tissue selectivity and biosafety, so it has a great application prospect. At present, there is very limited research on multi-electrode simulated ablation of histological electrical pulse. In this study, a circular multi-electrode ablation model of pulmonary vein will be built on COMSOL5.5 platform for simulation research. The results show that when the voltage amplitude reaches about 900 V, it can make some positions achieve transmural ablation, and the depth of continuous ablation area formed can reach 3 mm when the voltage amplitude reaches 1 200 V. When the distance between catheter electrode and myocardial tissue is increased to 2 mm, a voltage of at least 2 000 V is required to make the depth of continuous ablation area reach 3 mm. Through the simulation of electric pulse ablation with ring electrode, the research results of this project can provide reference for the voltage selection in the clinical application of electric pulse ablation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.