Abstract

A topology of high-frequency field intensity in the work area of a saddle-shaped coil for a nuclear qaudrupole resonance spectrometer was studied. With a view to determine magnetic field topology, a computational domain was created which is a model of a saddle-shaped coil physical structure. Finite element method was used to perform numerical simulation in COMSOL Multiphysics software. According to the results of calculations performed and the field maps obtained, the relative volume of coil work area was determined which makes 28.12% of its full volume. For such a volume the recommended size of samples under study is 12× 18× 10mm3.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.