Abstract

BackgroundDue to the heterogeneity of patient’s individual respiratory motion pattern in lung stereotactic body radiotherapy (SBRT), treatment planning dose assessment using a traditional four-dimensional computed tomography (4DCT_traditional) images based on a uniform breathing curve may not represent the true treatment dose delivered to the patient. The purpose of this study was to evaluate the accumulated dose discrepancy between based on the 4DCT_traditional and true 4DCT (4DCT_true) that incorporated with the patient’s real entire breathing motion. The study also explored a novel 4D robust planning strategy to compensate for such heterogeneity respiratory motion uncertainties.MethodsSimulated and measured patient specific breathing curves were used to generate 4D targets motion CT images. Volumetric-modulated arc therapy (VMAT) was planned using two arcs. Accumulated dose was obtained by recalculating the plan dose on each individual phase image and then deformed the dose from each phase image to the reference image. The “4 D dose” (D4D) and “true dose” (Dtrue) were the accumulated dose based on the 4DCT_traditional and 4DCT_true respectively. The average worse case dose discrepancy ( overline{Delta D} ) between D4D and Dtrue in all treatment fraction was calculated to evaluate dosimetric /planning parameters and correlate them with the heterogeneity of respiratory-induced motion patterns. A novel 4D robust optimization strategy for VMAT (4D Ro-VMAT) based on the probability density function(pdf) of breathing curve was proposed to improve the target coverage in the presence of heterogeneity respiratory motion. The data were assessed with a paired t-tests.ResultsWith increasing breathing amplitude from 5 to 20 mm, target overline{Delta {D}_{99}} , overline{Delta {D}_{95}} increased from 1.59,1.39 to 10.15%,8.66% respectively. When the standard deviation of breathing amplitude increased from 15 to 35% of the mean amplitude, overline{Delta {D}_{99}} , overline{Delta {D}_{95}} increased from 4.06,3.48 to 10.25%,6.63% respectively. The 4D Ro-VMAT plan significantly improve the target dose compared to VMAT plan.ConclusionWhen the breathing curve amplitude is more than 10 mm and standard deviation of amplitude is higher than 25% of mean amplitude, special care is needed to choose an appropriated dose accumulation approach to evaluate lung SBRT plan target coverage robustness. The proposed 4D Ro_VMAT strategy based on the pdf of patient specific breathing curve could effectively compensate such uncertainties.

Highlights

  • Due to the heterogeneity of patient’s individual respiratory motion pattern in lung stereotactic body radiotherapy (SBRT), treatment planning dose assessment using a traditional four-dimensional computed tomography (4DCT_traditional) images based on a uniform breathing curve may not represent the true treatment dose delivered to the patient

  • Digital lung cancer phantom simulation and dose accumulation The heterogeneous motion pattern mimicking real patient breathing curve was simulated, an example in Fig. 3a and b displayed the irregular respiratory motion simulated with different mean amplitude and different standard deviation of amplitude respectively

  • The results show that the dose degradation was sensitive to the offset from the reference, but the chance of exceeds the 4DCT_traditional average amplitude may varies depending on each specific patient

Read more

Summary

Introduction

Due to the heterogeneity of patient’s individual respiratory motion pattern in lung stereotactic body radiotherapy (SBRT), treatment planning dose assessment using a traditional four-dimensional computed tomography (4DCT_traditional) images based on a uniform breathing curve may not represent the true treatment dose delivered to the patient. The traditional 4DCT(4DCT_tradiational) is reconstructed by using 10 phase images to represent a periodic motion which may come from any single breathing cycle and may not fully represent true tumor position and motion [10, 11]. Such uncertainties will introduce variations of the dose in the target in the lung patient with irregular breathing curve/who cannot breathe homogeneously

Objectives
Methods
Results
Discussion
Conclusion
Full Text