Abstract
The Monte–Carlo (MC) program EGSnrc was used to establish the model of low-energy radiation field and low-energy free air ionisation chamber. The electron loss correction factor and the fluorescence scattering correction factor of the molybdenum-filtered X-ray radiation of 23 ∼ 35kv tungsten target were simulated. The results show that due to the low effective energy of the radiation and the energy is close, the influence of the electronic loss correction factor is negligible, and the fluorescence and scattering correction factors increase slightly with the increase of the effective energy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.