Abstract

We present a general framework of simultaneous needle shape reconstruction and control input generation for robot-assisted spinal injection procedures, without continuous imaging feedback. System input-output mapping is generated with a real-time needle-tissue interaction simulation, and single-core FBG sensor readings are used as local needle shape feedback within the same simulation framework. FBG wavelength shifts due to temperature variation is removed by exploiting redundancy in fiber arrangement. Targeting experiments performed on both plastisol lumbar phantoms as well as an ex vivo porcine lumbar section achieved in-plane tip errors of and , and total tip errors of and for the two testing environments. Our clinically inspired control strategy and workflow is self-contained and not dependent on the modality of imaging guidance. The generalizability of the proposed approach can be applied to other needle-based interventions where medical imaging cannot be reliably utilized as part of a closed-loop control system for needle guidance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.