Abstract

This paper presents comprehensive study on fat as propagation medium in abdominal implant communication system at low ultrawideband (UWB) frequency range 3.75-4.25 GHz. The main aim is to investigate how signal propagates through visceral and subcutaneous fat layers and how that information can be exploited in implant communication systems. The study is conducted using different methods: electromagnetic simulations, power flow analysis, propagation path calculations and radio channel measurements with animal meat pieces. Simulations are conducted using layer models and anatomical voxel models having different sizes. Results of channel simulations are verified with propagation path calculations. Power flow analysis on cross-cuts of the voxel models is conducted to investigate how the signal propagates inside the tissues. Furthermore, measurements using different animal meat pieces are performed to evaluate the impact of fat constitution on channel characteristics. It is found that similar tendency on fat propagation is seen in the evaluations with different methods. It is also observed that channel attenuation depends not only on the types and thicknesses of the tissues between transmitter and receiver antennas, but also how the tissues, especially fat, is located between the antennas. Channel attenuation difference between different voxels is maximum 14 dB in the studied antenna locations. Furthermore, propagation channel is evaluated with measurements using pork meat having different fat and muscle constitutions. It is found that antenna location respect to fat layers has clear impact on the channel strength although the fat tissue is not directly above the in-body antenna. The difference is noted to be 3-15 dB especially on the side peaks of channel impulse response. The knowledge on fat as a propagation medium is crucial when designing medical monitoring or implant communication systems. Location of antennas/sensor nodes for the monitoring devices can be established so that propagation through fat layer can be exploited.

Highlights

  • Medical implant communications and in-body propagation have been intensively studied topics for Wireless Body Area Network (WBAN) applications for several years [1]–[29]

  • The power flow depends on the antenna radiation pattern and remarkably on the tissues through which the signal travels

  • SUMMARY AND CONCLUSION This paper presented a study on the fat as propagation medium in WBAN abdominal implant communication systems

Read more

Summary

INTRODUCTION

Medical implant communications and in-body propagation have been intensively studied topics for Wireless Body Area Network (WBAN) applications for several years [1]–[29]. Similar power flow figures are presented in [22], but this study covers wider dB range and several different locations within the tissues to enable analysis of the propagation though the visceral fat as well. For the location ‘G’, the power loss is relatively high, 88 dB

CHANNEL EVALUATIONS WITH SIMULATIONS
CHANNEL EVALUATIONS WITH PORK MEAT MEASUREMENTS
Findings
DISCUSSION
VIII. DISCUSSION
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.