Abstract
Current guidelines for pediatric cardiopulmonary resuscitation suggest that supraglottic devices are alternatives for tracheal intubation with minimal interruption of chest compression. We examined the utility of three intubating supraglottic devices, air-Q® (air-Q), Ambu® aura-i (aura-i), and i-gel® (i-gel), utilizing manikin simulation. Twenty-two novice physicians performed securing of airway on an infant manikin with the three devices. We measured the rate of success on ventilation and the insertion time with or without chest compression. Successful ventilation rate did not significantly decrease with chest compression in the three devices (without chest compression: air-Q, 21/22; aura-i, 20/22; i-gel, 20/22, during chest compression: air-Q, 20/22; aura-i, 20/22; i-gel, 18/22). The insertion time with air-Q and aura-i did not extend significantly for chest compression. In contrast, the insertion time with i-gel was significantly extended in chest compression (P < 0.05). Air-Q and aura-i are more useful for airway management during chest compression than i-gel.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Pediatrics international : official journal of the Japan Pediatric Society
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.