Abstract

The artificial neural network algorithm has been used to simulate the response of an ionization chamber dosimetric system to the 137Cs irradiator facility at the National Institute for Standards, Egypt. The performance of the designed model has been investigated over different processes functions and a wide range of parameters till the optimum performance has been approached. The standard uncertainty of the designed model has been evaluated for different dose-rate levels. It has been found that the dose-rate can be evaluated using the designed model with an uncertainty < 2.0% for the dose-rate level $$<\,10^{2} \,{\text{mGy/h}}$$ and with an uncertainty < 0.5% for the dose-rate level $$[10^{2} {-}10^{3} ]\,{\text{mGy/h}}$$ . The bias of the designed model has been evaluated and compared with the usual interpolation algorithms. The errors in the evaluated dose-rate using these methods have exceeded the 100% error level and approached the 20% error level for the dose rates less and greater than $$10\, {\text{mGy/h}}$$ , respectively. While, the error in the evaluated dose-rate using the designed model did not exceed 7% and 2% for the same dose-rate levels, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.