Abstract

The need to monitor manual cleaning of high-risk endoscopes is recommended or more so required by the current endoscope reprocessing guidelines. The objective of this study was to establish the optimal extraction volume for colonoscopes and bronchoscopes and demonstrate the extraction efficacy for the ChannelCheck™ rapid test. The test soil utilized as a positive control was ATS2015 containing 20% defibrinated bovine blood. The extraction from the instrument channel of a colonoscope and bronchoscope was evaluated to establish the optimal extraction volume and the extraction efficacy for protein, carbohydrate and haemoglobin. Of the extraction volumes tested, 10mL was optimal for both colonoscopes and bronchoscopes. The extraction efficacy was 91% for carbohydrate, 83.7% for haemoglobin and 82.4% for protein. The limit of detection for these analytes by the ChannelCheck rapid test meet or exceed the established levels that correlate with adequate manual cleaning of flexible endoscopes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.